| Manufacturers | Radiation Oncology | Medical Physics | Diagnostic Imaging | Radiation Safety | Exercise Systems |
| MT-20100CF - Carbon Fiber Baseplate |
| Home : Med-Tec : Products : Radiation Oncology : Patient Positioning & Fixation : Head & Neck : Baseplates |
|
|
| Our rugged, radiotranslucent carbon fiber baseplate features indexing points for use with our IPPS indexing couch/Lok-Bar™, and is available in a standard supine model. The baseplate accommodates the Uni-frame™ and Drape and Mold thermoplastic-based head fixation systems. |
| Additional Components: | |
| Head & neck supports: | Thermoplastic masks: |
| Specifications: |
| All Baseplates are designed to provide linear indexing to within ± 1 mm and angular indexing (where applicable) to ± 2°, as referenced to the CT, simulator or treatment table when used in a typical clinical environment. However, the actual repeatability observed is dependent on the user’s skill, the actual clinical environment, and individual patient positioning variables. |
| Part #: | MT-20100CF | |
| Length: | 37.5 cm | |
| Width: | 25.4 cm | |
| Angular Tilt: | 0° (non-tilting) | |
| Max Support Weight1: | 20 kg max | |
| Useful life2 ([Typical use]/[Heavy use]) | 5 / 3 years | |
| Replaces (Old Model): | N/A | |
| Environment (Transportation/ Warehousing3,4,5) | - 20° Celsius to + 70° Celsius; 10 % to 85 % RH (non-condensing); Up to 3,000 m above MSL. | |
| Environment (Typical Usage): | + 15° Celsius to + 30° Celsius; 35 % to 65 % RH (non-condensing); Up to 1,000 m above MSL. | |
| Design And/Or Production Classifications/ Standards/ Certifications: | Class 1 MDD 93/42/EEC; EN46001; IEC 601; ISO 9001 and ISO 13485. | |
| Conformance Markings: | "CE" marked. | |
| 1. Baseplates are designed to support only the weight of the patient’s head, not the full weight of the patient. Exceeding the recommended support weight may cause product damage and/or injury to patient. |
| 2. Discontinue use if baseplate material shows signs of yellowing, or any discoloration, or has become translucent. |
| 3. Product must be packaged in undamaged, original packaging. |
| 4. Packaged product must not be exposed to sources of direct heat or sunlight; nor be exposed to more than a total maximum of 15 weeks at this environment. |
| 5. Allow product to acclimatize to the Typical Usage Environment for at least 24 hours before use. |
|
|