Collecting Volume: |
0.94 cm3 |
|
Nominal Calibration Factor: |
3.4 R/nC |
|
Centroid of Collecting Volume: |
1.5 mm from surface of window |
|
Collector Diameter: |
20.0 mm |
|
Window-Collector Gap: |
3.0 mm |
|
Window: |
Conductive Kapton film, 3.86 mg/cm2 |
|
Body, Collector, and Guard Material: |
Model A11TW – Shonka air-equivalent plastic C552
Model P11TW – Polystyrene equivalent plastic D400
Model T11TW – Shonka tissue-equivalent plastic A150
|
|
Maximum Polarizing Potential: |
Greater than 1000 V |
|
Inherent Leakage Currents:
|
Less than 10-15 Amperes |
|
Cable: |
50 Ohms, 29pF/f, 3 m long |
|
Signal Connector: |
Choice of coaxial BNC and triaxial BNC and TNC plugs and jacks. |
|
High Voltage Connector: |
Integral with triaxial connectors (shell of chamber is common with connector body). Banana, MHV, and SHV plugs are also available. |
|
Vent Tubes: |
Two 21 Ga stainless steel tubes which may be extended to connector via 22 Ga TFE tubing for possible immersion in liquid phantoms. (Chamber immersion requires a cover over the window.) |
|
Stem: |
7/16" OD black phenolic two-piece (4" + 5") |